Mechanical thrombectomy of major and massive pulmonary embolism with use of the Amplatz thrombectomy device. Academic Article uri icon

Overview

abstract

  • RATIONALE AND OBJECTIVES: To evaluate the feasibility of mechanical thrombectomy with the Amplatz thrombectomy device (ATD) in restoring patency of acutely thrombosed pulmonary arteries resulting from pulmonary embolism for the improvement of patient outcome. METHODS: Mechanical thrombectomy with the ATD (8F) was performed in nine consecutive patients with angiographically documented thrombus in the left or right pulmonary artery resulting from deep vein thrombosis (n = 4) or unknown cause (n = 5). RESULTS: The Miller index decreased from 18 to 11. In all patients, the majority of the thrombus in the pulmonary artery was cleared after a mean activation time of the ATD of 367 seconds. Thrombectomy was performed with the ATD alone (n = 4) or with additional long-term fibrinolysis therapy (n = 5) with infusion of recombinant tissue-type plasminogen activator. Pulmonary arterial pressure decreased from a mean of 57 mm Hg before mechanical thrombectomy to 55 mm Hg directly after the procedure and to 39 mm Hg after termination of the recombinant tissue-type plasminogen activator infusion. CONCLUSIONS: Mechanical thrombectomy with the ATD in patients with minor and major pulmonary embolism is technically feasible and safe. It is a potential alternative to drug-mediated thrombolysis and surgery. However, the incremental benefit of the ATD over conventional treatments could be shown only in a randomized controlled study.

publication date

  • June 1, 2001

Research

keywords

  • Pulmonary Embolism
  • Thrombectomy

Identity

Scopus Document Identifier

  • 0034991634

PubMed ID

  • 11410751

Additional Document Info

volume

  • 36

issue

  • 6