The Genesis stent: A new low-profile stent for use in infants, children, and adults with congenital heart disease. Academic Article uri icon

Overview

abstract

  • Placement of intravascular stents that can reach adult size in infants and smaller children has been limited by the large profile and poor flexibility of currently available stents. In vitro and in vivo testing of the Genesis stent was performed to evaluate crimpability, predeployment flexibility, and radial strength. Comparisons were made to the Palmaz iliac and IntraStent (IS) LD stents. Nine physicians placed 30 Genesis stents in swine pulmonary and systemic arteries to evaluate stent deliverability/crimpability. Two swine were recovered and underwent a second catheterization 8 weeks later, where the stents (n = 8) were reexpanded to maximal size. Angiographic and intravascular ultrasound (IVUS) assessments were performed. In vitro testing revealed the Genesis stent to have superior crimpability, flexibility, and comparable radial strength to the Palmaz iliac stent, and superior crimpability and radial strength and comparable flexibility to the IS LD series. During in vivo testing, the physicians graded the Genesis stent superior to the Palmaz stent regarding crimpability and deliverability, and superior to the IS LD stent in regard to crimpability, and comparable to or superior in deliverability. In the chronic animals, the Genesis stent was expanded up to maximal diameter 8 weeks following implantation. Angiographic and IVUS revealed no fractures no in-stent restenosis. The Genesis stent can be easily delivered through smaller sheaths, which will facilitate their use in infants and smaller children with vascular stenosis.

publication date

  • July 1, 2003

Research

keywords

  • Angioplasty, Balloon, Coronary
  • Heart Defects, Congenital
  • Prosthesis Design
  • Stents

Identity

Scopus Document Identifier

  • 0037678440

PubMed ID

  • 12822172

Additional Document Info

volume

  • 59

issue

  • 3