Improved magnetization preparation for navigator steady-state free precession 3D coronary MR angiography.
Academic Article
Overview
abstract
The purpose of this work was to investigate a new magnetization preparation scheme for navigator steady-state free precession (SSFP) 3D coronary MR angiography (MRA) that executes the navigator and fat saturation pulses in steady state after the dummy RFs in order to minimize the delay between the magnetization preparation and the image echoes. Compared to the previous preparation scheme that executes the navigator and fat saturation pulses before the dummy RFs, the new scheme was found to provide more effective motion suppression, significantly improved blood-to-myocardium contrast-to-noise ratio (46%, P < 0.001) at slightly but insignificantly decreased blood signal-to-noise ratio (SNR) (2%, P = 0.73), significantly reduced fat SNR (32%, P < 0.001), and better overall image quality (P = 0.05; Wilcoxon paired sample signed rank test).