Calibrated current divider network for precision current delivery during high-voltage transthoracic defibrillation. Academic Article uri icon

Overview

abstract

  • The design of a calibrated resistive-network current divider for precision current delivery during transthoracic defibrillation shocks is presented together with test results. The current divider presents a constant 50-ohm load to the defibrillator and thus maintains a constant pulse shape. Current is selected before the shock by setting three rheostats using a computer-generated calibration table. Following each shock, the data acquisition and display software updates the calibration table based on the measured value of transthoracic resistance. Over a range of 15-27 A, the root-mean-square (rms) error for delivered versus selected current was 0.48% for a 45-ohm resistive load, and 0.71% for a 100-ohm load. These test results were confirmed by animal experiments. Over 3 dogs, the rms error was 0.49% from 15-27 A and not greater than 1.5% over the entire 8-44 A range.

publication date

  • November 1, 2005

Research

keywords

  • Defibrillators
  • Electronics, Medical
  • Energy Transfer

Identity

Scopus Document Identifier

  • 27644530005

PubMed ID

  • 16285402

Additional Document Info

volume

  • 52

issue

  • 11