Dynamics of an intervertebral disc prosthesis in human cadaveric spines. Academic Article uri icon

Overview

abstract

  • Low-back pain is a common, disabling medical condition, and one of the major causes is disc degeneration. Total disc replacements are intended to treat back pain by restoring disc height and re-establishing functional motion and stability at the index level. The objective of this study was to determine the effect on range of motion (ROM) and stiffness after implantation of the ProDisc-L device in comparison to the intact state. Twelve L5-S1 lumbar spine segments were tested in flexion/extension, lateral bending, and axial rotation with axial compressive loads of 600 N and 1,200 N. Specimens were tested in the intact state and after implantation with the ProDisc-L device. ROM was not significantly different in the implanted spines when compared to their intact state in flexion/extension and axial rotation but increased in lateral bending. Increased compressive load did not affect ROM in flexion/extension or axial rotation but did result in decreased ROM in lateral bending and increased stiffness in both intact and implanted spine segments. The ProDisc-L successfully restored or maintained normal spine segment motion.

publication date

  • September 1, 2007

Identity

PubMed Central ID

  • PMC2504266

Scopus Document Identifier

  • 34848917831

Digital Object Identifier (DOI)

  • 10.1007/s11420-007-9049-0

PubMed ID

  • 18751788

Additional Document Info

volume

  • 3

issue

  • 2