Direct measurement of proximal isovelocity surface area by real-time three-dimensional color Doppler for quantitation of aortic regurgitant volume: an in vitro validation. Academic Article uri icon

Overview

abstract

  • OBJECTIVE: The proximal isovelocity surface area (PISA) method is useful in the quantitation of aortic regurgitation (AR). We hypothesized that actual measurement of PISA provided with real-time 3-dimensional (3D) color Doppler yields more accurate regurgitant volumes than those estimated by 2-dimensional (2D) color Doppler PISA. METHODS: We developed a pulsatile flow model for AR with an imaging chamber in which interchangeable regurgitant orifices with defined shapes and areas were incorporated. An ultrasonic flow meter was used to calculate the reference regurgitant volumes. A total of 29 different flow conditions for 5 orifices with different shapes were tested at a rate of 72 beats/min. 2D PISA was calculated as 2pi r(2), and 3D PISA was measured from 8 equidistant radial planes of the 3D PISA. Regurgitant volume was derived as PISA x aliasing velocity x time velocity integral of AR/peak AR velocity. RESULTS: Regurgitant volumes by flow meter ranged between 12.6 and 30.6 mL/beat (mean 21.4 +/- 5.5 mL/beat). Regurgitant volumes estimated by 2D PISA correlated well with volumes measured by flow meter (r = 0.69); however, a significant underestimation was observed (y = 0.5x + 0.6). Correlation with flow meter volumes was stronger for 3D PISA-derived regurgitant volumes (r = 0.83); significantly less underestimation of regurgitant volumes was seen, with a regression line close to identity (y = 0.9x + 3.9). CONCLUSION: Direct measurement of PISA is feasible, without geometric assumptions, using real-time 3D color Doppler. Calculation of aortic regurgitant volumes with 3D color Doppler using this methodology is more accurate than conventional 2D method with hemispheric PISA assumption.

publication date

  • January 24, 2009

Research

keywords

  • Aortic Valve Insufficiency
  • Blood Flow Velocity
  • Echocardiography, Doppler, Color
  • Echocardiography, Three-Dimensional
  • Image Interpretation, Computer-Assisted

Identity

PubMed Central ID

  • PMC3348870

Scopus Document Identifier

  • 60849096364

Digital Object Identifier (DOI)

  • 10.1016/j.echo.2008.11.031

PubMed ID

  • 19168322

Additional Document Info

volume

  • 22

issue

  • 3