PET/CT-guided interventions: personnel radiation dose. Academic Article uri icon

Overview

abstract

  • PURPOSE: To quantify radiation exposure to the primary operator and staff during PET/CT-guided interventional procedures. METHODS: In this prospective study, 12 patients underwent PET/CT-guided interventions over a 6 month period. Radiation exposure was measured for the primary operator, the radiology technologist, and the nurse anesthetist by means of optically stimulated luminescence dosimeters. Radiation exposure was correlated with the procedure time and the use of in-room image guidance (CT fluoroscopy or ultrasound). RESULTS: The median effective dose was 0.02 (range 0-0.13) mSv for the primary operator, 0.01 (range 0-0.05) mSv for the nurse anesthetist, and 0.02 (range 0-0.05) mSv for the radiology technologist. The median extremity dose equivalent for the operator was 0.05 (range 0-0.62) mSv. Radiation exposure correlated with procedure duration and with the use of in-room image guidance. The median operator effective dose for the procedure was 0.015 mSv when conventional biopsy mode CT was used, compared to 0.06 mSv for in-room image guidance, although this did not achieve statistical significance as a result of the small sample size (p = 0.06). CONCLUSION: The operator dose from PET/CT-guided procedures is not significantly different than typical doses from fluoroscopically guided procedures. The major determinant of radiation exposure to the operator from PET/CT-guided interventional procedures is time spent in close proximity to the patient.

publication date

  • November 15, 2012

Research

keywords

  • Laboratory Personnel
  • Occupational Exposure
  • Positron-Emission Tomography
  • Radiation Dosage
  • Radiation Monitoring
  • Technology, Radiologic
  • Tomography, X-Ray Computed

Identity

Scopus Document Identifier

  • 84879911419

Digital Object Identifier (DOI)

  • 10.1007/s00270-012-0515-9

PubMed ID

  • 23229891

Additional Document Info

volume

  • 36

issue

  • 4