  • PURPOSE: Evaluating the usefulness of diffusion-weighted spatio-temporal encoding (SPEN) methods to provide quantitative apparent diffusion coefficient (ADC)-based characterizations of healthy and malignant human breast tissues, in comparison with results obtained using techniques based on spin-echo echo planar imaging (SE-EPI). METHODS: Twelve healthy volunteers and six breast cancer patients were scanned at 3T using scanner-supplied diffusion-weighted imaging EPI sequences, as well as two fully refocused SPEN variants programmed in-house. Suitable codes were written to process the data, including calculations of the actual b-values and retrieval of the ADC maps. RESULTS: Systematically better images were afforded by the SPEN scans, with negligible geometrical distortions and markedly weaker ghosting artifacts arising from either fat tissues or from strongly emitting areas such as cysts. SPEN-derived images provided improved characterizations of the fibroglandular tissues and of the lesions' contours. When translated into the calculation of the ADC maps, there were no significant differences between the mean ADCs derived from SPEN and SE-EPI: if reliable images were available, both techniques showed that ADCs decreased by nearly two-fold in the malignant lesion areas. CONCLUSION: SPEN-based sequences yielded diffusion-weighted breast images with minimal artifacts and distortions, enabling the calculation of improved ADC maps and the identification of decreased ADCs in malignant regions.

publication date

  • July 12, 2014

Research

keywords

  • Breast
  • Breast Neoplasms
  • Diffusion Magnetic Resonance Imaging

Identity

Scopus Document Identifier

  • 84929653082

Digital Object Identifier (DOI)

  • 10.1002/mrm.25344

PubMed ID

  • 25045867

Additional Document Info

volume

  • 73

issue

  • 6