Functional validation of an implantable medical dosing device by MRI at 3T. Academic Article uri icon

Overview

abstract

  • Sustained release of a small molecule from a prototype implantable drug delivery device was monitored via MRI in an ex vivo tissue phantom over a period of two days. T1 mapping was used as a method to quantify analyte concentration. Continuous, controlled release was observed. The MRI methodology was thus found to be appropriate for device validation and quality assurance/control.

publication date

  • January 1, 2014

Research

keywords

  • Infusion Pumps, Implantable

Identity

Scopus Document Identifier

  • 84929494002

Digital Object Identifier (DOI)

  • 10.1109/EMBC.2014.6944091

PubMed ID

  • 25570459

Additional Document Info

volume

  • 2014