Development of a solvent-free analytical method for paracetamol quantitative determination in Blood Brain Barrier in vitro model. Academic Article uri icon

Overview

abstract

  • A Reversed Phase-High Performance Liquid Chromatography/Diode Array Detection method was developed and validated for paracetamol quantification in cell culture fluid from an in vitro Blood Brain Barrier model. The chromatographic method and sample preparation were developed using only aqueous solvents. The column was a XTerra RP18 150 × 4.6mm, 3.5 μm with a guard column XTerra RP18 20 × 4.6 mm, 3.5 μm at 35 °C and the mobile phase was composed by 100% formate buffer 20 mM at pH 4 and flow rate was set at 1 mL/min. The detection was at 242 nm. The sample was injected at 10 μL. Validation was performed using the accuracy profile approach. The analytical procedure was validated with the acceptance limits at ± 10% over a range of concentration from 1 to 58 mg L(-1). The procedure was then used in routine to determine paracetamol concentration in a brain blood barrier in vitro model. Application of the Unither paracetamol formulation in Blood Brain Barrier model allowed the determination and comparison of the transcellular passage of paracetamol at 37 °C and 4 °C, that excludes paracellular or non specific leakage.

publication date

  • February 21, 2015

Research

keywords

  • Acetaminophen
  • Blood-Brain Barrier
  • Chromatography, High Pressure Liquid
  • Chromatography, Reverse-Phase

Identity

Scopus Document Identifier

  • 84923346949

Digital Object Identifier (DOI)

  • 10.1016/j.jchromb.2015.02.019

PubMed ID

  • 25734939

Additional Document Info

volume