Device success and 30-day clinical outcome in patients undergoing preimplant valvuloplasty in transfemoral versus omitting valvuloplasty in transapical transcatheter aortic valve replacement. Academic Article uri icon

Overview

abstract

  • OBJECTIVES: The study objective was to evaluate the impact of preimplantation balloon valvuloplasty on procedural outcomes in high-risk or no-option patients with aortic stenosis undergoing Edwards Lifesciences (Irvine, Calif) Sapien valve placement. Paravalvular aortic regurgitation has been associated with long-term mortality after transcatheter aortic valve replacement. Whether omitting preimplant balloon valvuloplasty affects paravalvular aortic regurgitation after Edwards Sapien transcatheter aortic valve replacement is currently unknown. METHODS: We retrospectively analyzed the clinical outcome of 121 consecutive patients undergoing transapical (N = 50) or transfemoral (N = 71) Edwards Sapien transcatheter aortic valve replacement. Routinely, no preimplant balloon valvuloplasty was performed in transapical procedure as opposed to uniform preimplant balloon valvuloplasty in transfemoral cases. The incidence and severity of total and paravalvular aortic regurgitation and 30-day clinical outcomes were compared between the 2 cohorts. RESULTS: The average patient's age was 84.4 years, with a higher prevalence of smoking history (68% vs 42%, P = .005) and peripheral vascular disease (38% vs 20%, P = .03) in the patients undergoing transapical replacement. The preprocedural transthoracic echocardiographic and computed tomography findings were similar between the 2 cohorts. After transcatheter aortic valve replacement, the incidence of mild to moderate total aortic regurgitation (42% transfemoral vs 38% transapical), paravalvular aortic regurgitation (39% transfemoral vs 30% transapical), device success (88.7% transfemoral vs 94.0% transapical), and 30-day composite end points (9.9% transfemoral vs 14.0% transapical) were comparable in both groups. Multivariate regression analysis revealed male gender (odds ratio, 2.7; 95% confidence interval, 1.18-6.35; P = .02) but not preimplant balloon valvuloplasty as an independent predictor for mild or greater total aortic regurgitation. CONCLUSIONS: Compared with transapical transcatheter aortic valve replacement without preimplant balloon valvuloplasty, preimplant balloon valvuloplasty before transfemoral transcatheter aortic valve replacement resulted in a similar degree of prosthesis-related regurgitation, device success, and 30-day composite safety outcomes.

publication date

  • July 26, 2015

Research

keywords

  • Aortic Valve
  • Aortic Valve Stenosis
  • Balloon Valvuloplasty
  • Cardiac Catheterization
  • Heart Valve Prosthesis Implantation

Identity

Scopus Document Identifier

  • 84948711692

Digital Object Identifier (DOI)

  • 10.1016/j.jtcvs.2015.07.050

PubMed ID

  • 26277463

Additional Document Info

volume

  • 150

issue

  • 5