Right mini-thoracotomy: an adjunct to left subcostal automatic implantable cardioverter defibrillator implantation. Academic Article uri icon

Overview

abstract

  • High defibrillation thresholds are encountered in some patients during implantation of the automatic implantable cardioverter defibrillator (AICD). In a consecutive series of 52 patients with AICD implants, 46 had allepicardial implants by left subcostal thoracotomy, and 8 of these patients had thresholds greater than 30 J. Placement of an apicolateral large left ventricular patch and an extrapericardial large right atrial patch electrode through a limited right thoracotomy produced substantial reduction in their thresholds. Especially in patients who have had previous intrapericardial operations, this technique provides a simple solution to the problem of high thresholds.

publication date

  • May 1, 1989

Research

keywords

  • Electric Countershock
  • Thoracotomy

Identity

Scopus Document Identifier

  • 0024413352

PubMed ID

  • 2730201

Additional Document Info

volume

  • 47

issue

  • 5