Regular chondrocyte spacing is a potential cause for coherent ultrasound backscatter in human articular cartilage. Academic Article uri icon

Overview

abstract

  • The potential of quantitative ultrasound (QUS) to assess the regular cellular spacing in the superficial cartilage zones was investigated experimentally and numerically. Nine osteochondral samples, extracted from two human cadaver knee joints, were measured using a 50-MHz ultrasound scanning device and evaluated using Mankin score. Simulated backscattered power spectra from models with an idealized cell alignment exhibited a pronounced frequency peak. From the peak, cell spacing in the range between 15 and 40 μm between cell layers was detected with an average error of 0.2 μm. The mean QUS-based cell spacing was 28.3 ± 5.3 μm. Strong correlation (R2 = 0.59, p ≤ 0.001) between spacing estimates from light microscopy (LM) and QUS was found for samples with Mankin score ≤3. For higher scores, QUS-based spacing was significantly higher (p ≤ 0.05) compared to LM-based spacing. QUS-based spacing estimates together with other QUS parameters may serve as future biomarkers for detecting early signs of osteoarthrosis.

publication date

  • May 1, 2017

Research

keywords

  • Cartilage, Articular
  • Chondrocytes
  • Microscopy, Acoustic
  • Osteoarthritis, Knee
  • Ultrasonic Waves
  • Ultrasonography

Identity

PubMed Central ID

  • PMC6909996

Scopus Document Identifier

  • 85019043455

Digital Object Identifier (DOI)

  • 10.1121/1.4979339

PubMed ID

  • 28599554

Additional Document Info

volume

  • 141

issue

  • 5