Hemodynamic Effects of Concomitant Mitral Valve Surgery and Left Ventricular Assist Device Implantation. Academic Article uri icon

Overview

abstract

  • There are conflicting data regarding whether concomitant mitral valve surgery (MVS) at left ventricular assist device (LVAD) implantation is beneficial. This study aimed to assess the hemodynamic effects of concomitant MVS. Of all 73 enrolled patients, 44 patients had undergone concomitant MVS and 29 patients had not. Before LVAD implantation, MVS group had higher pulmonary capillary wedge pressure (p = 0.04). After LVAD implantation, MVS group had higher mean pulmonary artery pressure and cardiac output (CO). During the hemodynamic ramp study, MVS group had steeper CO slopes (0.18 [0.13 0.28] vs. 0.15 [0.08, 0.20] L/min/step; p = 0.04) at incremental LVAD speed and achieved a higher CO at the optimized set speed (5.5 [4.7, 6.9] vs. 4.9 [4.0, 5.7] L/min; p = 0.03). One-year freedom from death or heart failure readmission was statistically comparable between the two groups (61% vs. 80%, p = 0.20). Thus far, after LVAD implantation and concomitant MVS, patients had increased pulmonary hypertension, despite having higher CO and a better response of CO at incremental LVAD speed. The implication of hemodynamic features after concomitant MVS on clinical outcomes warrants further investigation.

publication date

  • April 1, 2020

Research

keywords

  • Heart-Assist Devices
  • Hemodynamics
  • Mitral Valve

Identity

PubMed Central ID

  • PMC7081836

Scopus Document Identifier

  • 85082561210

Digital Object Identifier (DOI)

  • 10.1097/MAT.0000000000000999

PubMed ID

  • 30985304

Additional Document Info

volume

  • 66

issue

  • 4