The aiming device for cervical distractor pin insertion: a proof-of-concept, feasibility study.
Academic Article
Overview
abstract
BACKGROUND: Restoration of cervical lordosis after anterior discectomy and fusion is a desirable goal. Proper insertion of the vertebral distraction or Caspar pin can assist lordotic restoration by either putting the tips divergently or parallel to the index vertebral endplates. With inexperienced surgeons, the traditional free-hand technique for Caspar pin insertion may require multiple insertion attempts that may compromise the vertebral body and increase radiation exposure during pin localization. Our purpose is to perform a proof-of-concept, feasibility study to evaluate the effectiveness of a pin insertion aiming device for vertebral distraction pin insertion. METHODS: A Smith-Robinson approach and anterior cervical discectomy were performed from C3 to C7 in 10 human cadaveric specimens. Caspar pins were inserted using a novel pin insertion aiming device at C3-4, C4-5, C5-6, and C6-7. The angles between the cervical endplate slope and Caspar pin alignment were measured with lateral cervical imaging. RESULTS: The average Superior Endplate-to-Caspar Pin angle (SE-CP) and the average Inferior Endplate-to-Caspar Pin angle (IE-CP) were 6.2 ± 2.0° and 6.3 ± 2.2° respectively. For the proximal pins, the SE-CP and the IE-CP were 4.0 ± 1.1°and 5.2 ± 2.4° respectively. For the distal pins, the SE-CP and the IE-CP were 7.7 ± 1.4° and 6.2 ± 2.0° respectively. No cervical endplate violations occurred. CONCLUSION: The novel Caspar pin insertion aiming device can control the pin entry points and pin direction with the average SE-CP and average IE-CP of 6.2 ± 2.0° and 6.3 ± 2.2°, respectively. The study shows that the average different angles between the Caspar pin and cervical endplate are less than 7°.