Complication rates in concurrent inflatable penile prosthesis and incontinence surgery: Comparing the penoscrotal versus perineal incision approach.
Academic Article
Overview
abstract
The main objective of this study was to assess the IPP complication rates of patients undergoing placement via perineal incision versus more traditional penoscrotal approach in synchronous dual implantation. We identified 38 patients who underwent dual implantations of an IPP and AUS or urethral sling from 2011 to 2021 at a single tertiary center, 24 via perineal and 14 via penoscrotal incision. All IPP implants were done by a single surgeon. IPP postoperative complications were captured using the Clavien-Dindo classification at three separate time points, < 30 days, 30 days - 6 months, and > 6 months. The perineal group had two complications, IPP explantation due to rectourethral fistula (Grade III, > 6 months), and IPP explantation due to chronic genital pain (Grade III, > 6 months). The penoscrotal group had three complications, post-operative urinary retention requiring catheterization (Grade I, < 30 days), incision site infection (Grade I, < 30 days), and IPP explantation due to infection (Grade III, 30 days to < 6 months). There was no statistically significant difference in rate of patients with IPP complications between the two groups (p = 0.546) or in rate of IPP device malfunction (p = 0.264). These preliminary findings suggest that the single perineal incision is a viable surgical approach in synchronous dual implantation.