Quantitative velocity images from thick slab 2D phase contrast. Academic Article uri icon

Overview

abstract

  • Thick slab two-dimensional phase contrast (2D PC) angiography provides a large amount of anatomical information in a short acquisition time. Quantitative velocity information is, however, destroyed by the necessary projection dephasing gradient. We present a 2D PC acquisition scheme which retains quantitative velocity information in a thick slab acquisition. A thick slab acquisition produces an image which is a projection through an entire vessel. Accurate velocity measurements must take into account the intravoxel phase cancellation caused by a distribution of velocities within the vessel. In addition to the description of the data acquisition scheme, we outline a method for determining the mean velocity within a vessel having a laminar flow distribution. It is shown that without the consideration for intravoxel phase cancellation, the mean velocity measurement from thick slab images overestimates the true mean velocity by 10%.

publication date

  • February 1, 1993

Research

keywords

  • Blood Vessels
  • Magnetic Resonance Imaging

Identity

Scopus Document Identifier

  • 0027531942

Digital Object Identifier (DOI)

  • 10.1002/mrm.1910290210

PubMed ID

  • 8429786

Additional Document Info

volume

  • 29

issue

  • 2