Application of the continuity equation and valve resistance to the evaluation of St. Jude Medical prosthetic aortic valve dysfunction. Academic Article uri icon

Overview

abstract

  • Doppler echocardiography was applied to the assessment of patients with surgically documented St. Jude medical aortic valve dysfunction. Derivation of effective orifice area and Doppler velocity index with the continuity equation and calculation of valve resistance accurately differentiated stenotic from regurgitant and normal valves.

publication date

  • November 1, 1997

Research

keywords

  • Aortic Valve Insufficiency
  • Aortic Valve Stenosis
  • Echocardiography, Doppler
  • Heart Valve Prosthesis
  • Prosthesis Failure

Identity

Scopus Document Identifier

  • 0030667656

PubMed ID

  • 9359564

Additional Document Info

volume

  • 80

issue

  • 9