Knee kinematics in genesis total knee arthroplasty. A comparison of different tibial designs with and without posterior cruciate substitution in cadaveric specimens. Academic Article uri icon

Overview

abstract

  • Knee joint kinematics after total knee arthroplasty (TKA) are not well understood. This study measured knee kinematics before and after TKA in six cadaveric specimens. Different tibial surface contours (standard, flat, and dished) and slopes (10 degrees and 15 degrees) with the posterior cruciate ligament (PCL) intact as well as a posterior-stabilized design were studied. The anteroposterior and proximal-distal displacements of the tibia relative to the femur were measured during active knee extension. For the standard design, it was possible to restore the normal position of the tibia relative to the femur at 90 degrees of flexion to within 2 mm; however, restoration of the tibiofemoral position was not achieved with the knee in the extended position. At 90 degrees of flexion, all of the TKA components resulted in a posteriorly positioned tibia compared with the normal knee. The standard, flat, and dished components shifted the tibia into a proximal position compared with the normal knee, while the 15 degrees and posterior-stabilized components shifted the tibia distally. With the knee in the extended position, the standard, flat, dished, and posterior-stabilized designs placed the tibia posteriorly and proximally compared with the normal position of the tibia relative to the femur. The 10 degrees and 15 degrees sloped components placed the tibia in a more anterior position. Of the PCL-retaining designs tested, the 10 degrees sloped tibial component produced the closest to normal knee kinematics. Overall, normal kinematic behavior of the knee was not restored after TKA.

publication date

  • January 1, 1997

Research

keywords

  • Arthroplasty, Replacement, Knee
  • Knee Joint
  • Range of Motion, Articular

Identity

Scopus Document Identifier

  • 0031217655

PubMed ID

  • 9421596

Additional Document Info

volume

  • 10

issue

  • 4