Quantitation of renal perfusion using arterial spin labeling with FAIR-UFLARE. Academic Article uri icon

Overview

abstract

  • Quantitative perfusion imaging of human kidneys was performed using arterial spin labeling MRI with a fast spin echo readout-sequence. Perfusion maps of centrally located single slices were obtained in axial and coronal orientations. In ten healthy volunteers, the mean value of perfusion was 213+/-55 mL/(100g min) with a range from 140 to 319 mL/(100g min). These results are in accordance with literature data, considering the fact that FAIR only measures the perfusion component normal to the imaging plane. Intra-individual reproducibility errors of +/-11% were smaller than the natural interindividual variability of renal perfusion (SD = +/- 25%). Perfusion in the cortex was approximately 3-4 times higher compared to the medulla. Considering the relatively high resolution of 2x2x10 mm3, the ability to quantify perfusion, and the lack of ionizing radiation and contrast media, this technique should prove useful in diagnosing renal pathologies that are associated with reductions in tissue perfusion.