Radiation-force technique to monitor lesions during ultrasonic therapy. Academic Article uri icon

Overview

abstract

  • This report describes a monitoring technique for high-intensity focused ultrasound (US), or HIFU, lesions, including protein-denaturing lesions (PDLs) and those made for noninvasive cardiac therapy and tumor treatment in the eye, liver and other organs. Designed to sense the increased stiffness of a HIFU lesion, this technique uniquely utilizes the radiation force of the therapeutic US beam as an elastographic push to detect relative stiffness changes. Feasibility was demonstrated with computer simulations (treating acoustically induced displacements, concomitant heating, and US displacement-estimation algorithms) and pilot in vitro experimental studies, which agree qualitatively in differentiating HIFU lesions from normal tissue. Detectable motion can be induced by a single 5 ms push with temperatures well below those needed to form a lesion. Conversely, because the characteristic heat diffusion time is much longer than the characteristic relaxation time following a push, properly timed multiple therapy pulses will form lesions while providing precise control during therapy.

publication date

  • November 1, 2003

Research

keywords

  • Computer Simulation
  • Image Processing, Computer-Assisted
  • Ultrasonic Therapy