Effect of comfort pads and incubator design on neonatal radiography.
Academic Article
Overview
abstract
BACKGROUND: There has been increasing interest in patient dose reduction in neonatal intensive care units. Removing comfort pads for radiography has been identified as a potential means to decrease patient dose. OBJECTIVE: To assess the effect of comfort pads and support trays on detector entrance exposure (DEE) and image quality for neonatal radiography, and its implication for patient dose. MATERIALS AND METHODS: Comfort pads and support trays from three incubator and warmer systems were examined. The attenuation of the primary beam by these structures was measured using a narrow beam geometry. Their effect on DEE and image quality was then assessed using typical neonatal chest radiography techniques with three configurations: 1) both the comfort pad and support included in the beam, 2) only the support tray included and 3) both the comfort pad and support tray removed. RESULTS: Comfort pads and support trays were found to attenuate the primary beam by 6-15%. Eliminating these structures from the X-ray beam's path was found to increase the detector entrance exposure by 28-36% and increase contrast-to-noise ratio by more than 21%, suggesting room for patient dose reduction when the same image quality is maintained. CONCLUSION: Comfort pads and tray support devices can have a considerable effect on DEE and image quality, with large variations among different incubator designs. Positioning the image detector directly underneath neonatal patients for radiography is a potential means for patient dose reduction. However, such benefit should be weighed against the risks of moving the patient.