Comparison of Two High-Throughput Reverse Transcription-PCR Systems for the Detection of Severe Acute Respiratory Syndrome Coronavirus 2. Academic Article uri icon

Overview

abstract

  • Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has emerged as the cause of a worldwide pandemic. Many commercial SARS-CoV-2 reverse transcription-PCR (RT-PCR) assays have received Emergency Use Authorization from the U.S. Food and Drug Administration. However, there are limited data describing their performance, in particular the performance of high-throughput SARS-CoV-2 RT-PCR systems. We analyzed the diagnostic performance of two high-throughput systems: cobas 6800 and Panther Fusion, and their associated RT-PCR assays, with a collection of 389 nasopharyngeal specimens. The overall agreement between the platforms was 96.4% (375/389). Cohen's kappa analysis rated the strength of agreement between the two platforms as "almost perfect" (κ = 0.922; standard error, 0.051). Furthermore, there was no significant difference between corresponding cycle threshold values generated on the two systems (P value = 0.88; Student's t test). Taken together, these data imply that the two platforms can be considered comparable in terms of their clinical performance. We believe that this information will be useful for those who have already adopted these platforms or are seeking to implement high-throughput RT-PCR testing to stem the SARS-CoV-2 pandemic.

publication date

  • July 23, 2020

Research

keywords

  • Betacoronavirus
  • Coronavirus Infections
  • High-Throughput Screening Assays
  • Pneumonia, Viral
  • Reverse Transcriptase Polymerase Chain Reaction

Identity

PubMed Central ID

  • PMC7383551

Scopus Document Identifier

  • 85084617034

Digital Object Identifier (DOI)

  • 10.11613/BM.2012.031

PubMed ID

  • 32381643

Additional Document Info

volume

  • 58

issue

  • 8