Multicenter evaluation of the new VITEK 2 advanced colorimetric yeast identification card.
Academic Article
Overview
abstract
-
The performance of the new VITEK 2 Advanced Colorimetry yeast identification (YST) card for use with the VITEK 2 system (bioMérieux, Inc., Hazelwood, MO) was compared to that of the API 20C AUX (API) system (bioMérieux SA, Marcy-l'Etoile, France) in a multicenter evaluation. A total of 12 quality control, 64 challenge, and 623 clinical yeast isolates were used in the study. Comparisons of species identification, platform reliability, and substrate reproducibility were made between YST and API, with API considered the reference standard. Quality control testing to assess system and substrate reproducibility matched expected results >/=95% of the time. The YST card correctly identified 100% of the challenge strains, which covered the species range of the manufacturer's performance claims. Using clinical isolates, the YST card correctly identified 98.5%, with 1.0% of isolates incorrectly identified and 0.5% unidentified. Among clinical isolates, the YST card generated fewer low-discrimination results (18.9%) than did API (30.0%). The time to identification with YST was 18 h, compared to 48 to 72 h with API. The colorimetric YST card used with the VITEK 2 provides a highly automated, objective yeast identification method with excellent performance and reproducibility. We found this system useful for timely and accurate identification of significant yeast species in the clinical microbiology laboratory.
publication date
published in
Research
keywords
-
Colorimetry
-
Mycological Typing Techniques
-
Mycoses
-
Yeasts
Identity
PubMed Central ID
Scopus Document Identifier
PubMed ID
Additional Document Info
has global citation frequency
volume
issue